W203 2.0 Electrolytic Method Water Vapor Transmission Rate Tester
Product introduction
This product is based on the testing principle of electrolysis method and is developed, designed and manufactured in accordance with GB/T 21529, ISO15106-3 and other standards. It is used to test the water vapor transmission rate (volume) of samples under set temperature and humidity conditions. It is suitable for testing the water vapor transmission performance of films, sheets, paper, packaging and various materials in the fields of food, medicine, medical equipment, daily chemicals, photovoltaic electronics, etc. It is an ideal configuration instrument for offline or online testing of the barrier properties of packaging materials by production units in the packaging industry.
Test principle
W203 2.0 Water Vapor Permeability Analyzer uses the principle of electrolytic method. The pre-treated sample is fixed in the middle of the test chamber, and the test chamber is divided into high humidity side and low wet side. Compressed air flows on one side of the film and dry nitrogen (carrier gas) flows at a fixed flow rate on the other side. Due to the presence of a humidity gradient, water vapor penetrates from the high-humidity side to the low-humidity side, and the water vapor through the sample is carried to the electrolytic sensor by flowing dry nitrogen. Parameters such as the water vapor transmittance of the sample are obtained from the electrical signal output by the sensor.

Working principle diagram
Standard
ISO15106-3, ASTM E398-2013, DIN 53122-2, YBB00092003-2015, GB/T 21529-2008
Specification
| Item | Technical Parameters |
| Test Range | 0.001~40 g/(m²·24h) (Films and Sheets) |
| Resolution | 0.0001 g/(m²·24h) (Films and Sheets) |
| Temperature Control Range | 15~45°C |
| Temperature Control Accuracy | ±0.1°C |
| Humidity Control Range | (5~90)%RH |
| Humidity Control Accuracy | ±1%RH |
| Permeation Area | 50.24 cm² |
| Sample Size | Φ100 mm |
| Sample Thickness | ≤3 mm |
| Test Quantity | 3 pieces |
| Carrier Gas | 99.999% Nitrogen (user-supplied) |
| Carrier Gas Pressure | ≥0.1 MPa |
| Carrier Gas Flow Rate | 0~100 mL/min |
| Gas Source Interface | 1/8 inch Metal Tube |
| Overall Dimension | 670 mm × 550 mm × 400 mm |
| Weight | 57.1 kg |
| Power | 750 W |
| Power Supply | AC 220 V, 50 Hz |
Application
|
|
Film |
Water vapor transmittance test of various plastic films (PP/PET/PE/PVC/BOPP/CPP, etc.), plastic composite films, paper-plastic composite films, metal composite films, co-extrusion films, aluminized films, degradable packaging films (PLA/PBAT/PBS, etc.) and other film-like materials. |
|
|
Sheet |
Water vapor transmittance test of solid pharmaceutical hard sheets (PP/PVC/PTP, etc.), metal composite sheets, rubber sheets and other flakes. |
|
|
Paper, cardboard and its composites |
Water vapor transmittance test of coated paper, silicone paper, cigarette bag aluminized paper, paper aluminum-plastic composite sheet and other paper and cardboard. |
|
|
Medicinal patches |
Water vapor transmission performance test of medical plasters |
|
|
Package |
Customizable fixtures can be extended to packages such as pharmaceutical polyethylene bottles, sealed bags, pharmaceutical ointment tubes, infusion hoses, plastic trays, etc. |
Configuration
|
Standard configuration |
Sealing ring, syringe sealing ring, fork wrench, cross screwdriver, air pipe, conversion joint, voltage regulator, swimming separator, air compressor joint, mouse, wooden box, metal air pipe |
|
Optional |
Computer, calibration certificate |
|
Remark |
High purity over 99.999%, 1 bottle of nitrogen in a 40-liter steel cylinder. 2. Standard laboratory environment, With air conditioning, the temperature is 23℃±2℃. 3. Power requirements: 220 V regulated power supply, one socket with three holes and three switches. 4. Computer requirements: standard configuration (Windows 10, with a nine-pin serial port). 5. Requirements for experimenters: 1-3 test operators, technical secondary school or above, with certain computer knowledge. 6. Drying dish (all samples need to be dehydrated and degassed for 24 hours). 7. Air compressor (Pressure above 0.7 MPa) 8. Grounding requirements: the power supply is well grounded. |
Hot Tags: WVTR testing machine, China, suppliers, manufacturers, factory, custom, quotation, in stock, WVTR Permeability Analyzer, WVTR Testing Equipment for Barrier Food Packaging, Water Vapour Permeability Analyzer of Packaging Films, Water Vapor Transmission Tester of Laminated Films, Water Permeability Apparatus, WVTR Testing Machine