Key Specification | Specification Range | Resolution | Accuracy |
A Route Sampling Flow | 28.3L/min | 0.1L/min | Within ± 2.5% |
B Route Sampling Flow | 28.3L/min | 0.1L/min | Within ±2.5% |
Spray Flow | (8~10)L/min | 0.1L/min | Within ±2.5% |
Peristaltic Pump Flow | (0.006~3.0)mL/min | 0.001ml/min | Within ±2.5% |
Front Pressure of A Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.5% |
Front Pressure of B Route Sampling Flowmeter | (-20~0)kPa | 0.01kPa | Within ±2.5% |
Front Pressure of Spray Flowmeter | (0~300)kPa | 0.1kPa | Within ±2.5% |
Ambient Temperature | (-40~99)℃ | 0.1℃ | Within ±2.5% |
Negative Pressure of the Aerosol Chamber | (-90~-120)Pa | 0.1Pa | Within ±2.0% |
Negative Pressure of the Cabinet | -50~-200pa | ||
Data Storage Ability | >100000 sets | ||
Vortex Mixer Test Tube Specification and Quantity | Φ16×150mm test tube,8pcs | ||
High Efficiency Particulate Air Filter Properties | Filter Efficiency≥99.99% for particles> 0.3μm | ||
Median diameter of aerosol generator mass | Average diameter: (3.0±0.3)μm; Geometric Standard Deviation≤1.5 | ||
Double Routes 6-lever Andersen Sampling apparatus | Level1>7μm; Level 2: (4.7~7)μm; | ||
Size of Aerosol Chamber | 600×85×3mm (Length×Diameter×Thickness) | ||
Total number of positive quality control sampler particles | (2200±500)cfu | ||
Negative Chamber Flow Speed | ≥5m3/min | ||
Size of Negative Pressure Chamber Door(W×D) | 1000×730mm | ||
Size of Main Machine(W×D×H) | 1180×650×1300mm | ||
Size of Supporting Frame(W×D×H) | 1180×650×600mm, Height within 10cm is adjustable | ||
Noise of the Tester | <65dB(A) | ||
Weight | 150kg | ||
Power Supply | AC220V±10%, 50Hz | ||
Power Consumption | <1500W | ||
Features
1. Negative pressure experiment system to ensure the safety of operators;
2. Negative pressure cabinet with built-in peristaltic pump and A and B two-way six-stage Andersen;
3. The flow rate of the peristaltic pump can be set;
4. Dedicated microbial aerosol generator can set the volume of bacterial liquid spray flow, and the atomization effect is good;
5. Embedded high-speed industrial microcomputer control;
6.10.4-inch industrial-grade high-brightness color touch display;
7. USB interface, support U disk data transfer;
8. Built-in high-brightness lighting in the cabinet;
9. Built-in leakage protection switch to protect operator safety;
10. Insulation and flame retardant between inner and outer layers;
11. The front switch glass door is convenient for experimenters to observe and operate;
12. Detachable stand, the height of stand is adjustable;
13. Support and move dual-purpose casters.
Hot Tags: mask bacterial filtration efficiency (BFE) tester, China, suppliers, manufacturers, factory, custom, quotation, in stock, Face Mask Differential Pressure Tester, Exhalation Valve Air Tightness Tester, Mask Particle Filtration Efficiency Tester for Mask, Mask Bacterial Filtration Efficiency BFE Tester, Lab Synthetic Blood Penetration Non Woven Tester, Disposable Medical Mask Respiratory Resistance Tester